qPCR assay Triplicate Ct SD
0
0
Entering edit mode
2.0 years ago
Maria17 ▴ 40

Hi All,

Currently, I'm working on a qPCR assay. Since all targets are done in triplicate, I don't know the acceptable range of Ct Value Standard Deviation between replicates. Most of my targets have a CT SD less than 1, but some have a CT SD between 1-2. What should I do with those targets? Should I discard them or should I leave them as they are?

Any recommendations or solutions would be appreciated.

Thanks

Maria

miRNa qPCR RT-qPCR • 800 views
ADD COMMENT
0
Entering edit mode

There's no strict or generally accepted cutoff, but I probably wouldn't use this data for final reporting in a publication. I'd look into why this one target has an unusually high CT SD and work to alleviate that issue first. I suggest freezing the cDNA/RNA you're analyzing, as its probably some problem related to primer efficiency, reaction conditions, low starting copy number, etc.. which you could optimize and then re-run your initial targets with hopefully lower CT SD

ADD REPLY
0
Entering edit mode

Thank you for your response. How about the CT values for the two plasma samples (tubes) from the same patient?

In some cases, I have to measure the same miRNA target in several tubes from the same patient, but some of the target values are different. According to what I understand, the target CT values in both tubes from the same patient should be identical or similar.

It would be great to have your response to this issue.

Regards Maria

ADD REPLY

Login before adding your answer.

Traffic: 2499 users visited in the last hour
Help About
FAQ
Access RSS
API
Stats

Use of this site constitutes acceptance of our User Agreement and Privacy Policy.

Powered by the version 2.3.6